Original Article Relationship between the Measurement Values of the Digitial Dust Indicator and Personal Dust Sampler.
Young No Yoon, Jeong Joo Lee, Ho Keun Chung, Ho Chun Choi
Journal of Preventive Medicine and Public Health 1987;20(2):255-260
DOI: https://doi.org/
  • 1,968 Views
  • 19 Download
  • 0 Crossref
  • 0 Scopus

The digital dust indicator (Sibata P-5), one of the direct-reading instruments was evaluated for the respirable dust in the underground coal mine environments. As a reference, respirable dust was determined using three cyclones and/or impactors. All the tests were performed on aerosol in twenty underground coal mines. The coefficients of mass-relative concentration were 0.067+/-0.054 (Mean+/-Standard deviation) (range: 0.006-0.172). The relationship between relative concentration and temperature was not significant statistically. Also, the relationship of relative concentration and relative humidity was not significant. Mass concentration and relative concentration were 5.31+/-5.22 mg/m3 and 162+/-163 CPM (Mean+/-Standard deviation) respectively. The range of mass concentration was 1.22-22.69 mg/m3; relative concentration 16-628 CPM. The relationship of mass concentration and relative concentration was not significant in these ranges.